Machine QA Patient QA Diagnostic Health Physics
CDX 2000B Electrometer Low Noise, Repeatable Measurements
The CDX 2000B is a stable, lightweight electrometer for use in external beam dosimetry measurements and precise HDR brachytherapy calibrations.
Dimensions
-
Height7 cm (2.75 in)
-
Width22.2 cm (8.24 in)
-
Length23 cm (9 in)
-
Weight1.4 kg (3.00 lbs)
Three Modes
-
Rate0.001 nA – 500.00 nA, 1 pA resolution
-
Charge0.01 nC – 999,999 nC, 10 pC resolution
-
CombinedAccumulated charge and the current readings are displayed simultaneously
Charge Collections
-
TimedUser set duration (Range: 0 – 600 sec; Increment: 15 sec; Resolution: 1 sec)
-
ContinuousUnlimited duration with manual stop
Bias Voltage
-
Nominal ± 450 volt bias
-
5 User Bias Setting-450, - 300, - 150, 0; 150, 300, 450 (VDC)
-
Accuracy± 0.3 volt
Specifications
-
Repeatability± 0.1%
-
Long-Term Stability0.5% (over one year)
-
Stabilization Time± 0.1%
-
Response Time< 3 s
-
Non-Linearity± 0.25%
-
ConformityCE 93/42/EEC
-
Zero Drift± 0.25%
-
Zero Shift± 0.25%
-
DisplayBacklit LCD, 2 x 20 with 5/16" characters
-
InputBNC two lug, triaxial connector
-
ZeroingAutomatic zero function, user activated
-
OutputIsolated RS-232; unidirectional 19,200 baud rate; 8 data bits; no parity; 1 stop bit
-
Power100-240 VAC, 0.5 A max, 50/60 Hz input to external power supply, 9 VDC, 1.7 A power supply output to electrometer input, UL/TUL listed power supply; internal battery: 8 hrs per charge
-
Product StandardsCE 93/42/EEC