Pro CT Dose Phantom Measure Radiation Dose Output for CT and CBCT Scanners

The Pro-CT Dose Phantom is an exceptionally accurate phantom for measuring radiation dose output for CT and CBCT scanners simulating adult and pediatric head and body scans.

Pro CT Dose Phantom part of complete solution for: Diagnostic CT QA

Pro CT Dose Phantom Overview

The Pro-CT Dose Phantom is an exceptionally accurate phantom for measuring radiation dose output for CT and CBCT scanners simulating adult and pediatric head and body scans. The nested phantom adjusts into 3 sizes depending on protocol and is suitable for CTDI measurements essential within your CT/CBCT quality control program.

Includes detector cavity for CT Ion Chamber and 13 acrylic rods for plugging all phantom holes. Works in accordance with FDA performance standard for diagnostic X-ray systems (21CFR 1020.33) and CTDI measurements according to TG 66, IEC 60601-2-44, IEC 61223-3-5, IEC 61223-2-6.

  • Pro-CT Dose L (adult body) - 150 mm cylinder, diameter: 320 mm.
  • Pro-CT Dose M (adult head / pediatric body) - 150 mm cylinder, diameter: 160 mm,
  • Pro-CT Dose S (pediatric head) - 150 mm cylinder, diameter: 100 mm.
  • Each phantom includes five Ø13.1 mm CT probe holes (one in the middle and 4 around the perimeter, 90 ° apart and 10 mm from the edge. Acrylic rods for plugging all included.
  • Smaller cylinders can be inserted into larger ones together creating one solid cylinder
  • Engraved crosshair markings on phantoms for easier positioning and beautiful polished finish
  • Includes portable carrying case.