SuperMAX™ Electrometer
The premier two-channel, reference-grade electrometer on the market.
SuperMAX™ Electrometer Overview
Two Independent Measurement Channels
- Independent control over range, bias voltage, and applied system factor.
- Both channels feature extensive range (0.001 pA to 500.0 nA, 0.001 pC to 999.9 µC) for use in a variety of applications, including ratio measurements.
Built-In Detector Library
- Eliminates the need for a PC or to apply system factor and temperature/pressure corrections.
- Store over 100 detectors and calibrations.
- Apply individual factors to one or both measurement channels.
Easy-to-Use Interface
- Large, color display with built-in touchscreen.
- On-screen keypad and pull down menus allows for easy access to settings and detector library.
Store Data Collection
- Measurements are stored from all data collection.
- Set automatic repeat measurements.
Exradin W1 Scintillator Integration
Two dedicated modes in the SuperMAX allow for seamless calculations using the Exradin W1 Scintillator.
Rate Mode Display Ranges
-
LOW: 0.001 pA – 500.0 pA
-
HIGH: 0.001 nA – 500.0 nA
Charge Mode Display Ranges
-
LOW: 0.001 pC – 999.9 μC
-
HIGH: 0.001 nC - 999.9 μC
Rated Effective Range on Input Currents Per IEC 60731 for Reference Grade
-
0.400 pA – 500.0 nA (± .25% of minimum effective scale reading)
Measuring Range
-
LOW: 0.400 pA - 500 pA, 0.400 pC - 999.9 μC
-
HIGH: 0.400 nA - 500 nA, 0.400 nC - 999.9 μC
Bias Voltage
-
Nominal ±1000 V
-
0, (±) 100 - 1000 V in 1 V increments
Range Switching
-
User selectable
Display Resolution
-
LOW: Rate mode/Charge mode0.001 pA / 0.001 pC
-
HIGH: Rate mode/Charge mode0.001 nA / 0.001 nC
Repeatability
-
Low range: 0.25%
-
High range: ± 0.1%
Long Term Stability
-
± 0.5% over 1 year
Stabilization Time
-
± 0.1% of value at 1 hour for measurements taken at 15 minutes and 6 hours
Non-linearity:
-
± 0.25%
Zero Drift
-
Rate mode: ± 0.25% of minimum effective scale reading (LOW/HIGH: < ± 0.001 pA / < ± 0.001 nA)
Zero Shift
-
Rate mode: ± 0.25% of minimum effective scale reading (LOW/HIGH: < ± 0.001 pA / < ± 0.001 nA)
Charge Leakage Current
-
LOW/HIGH: < ± 0.001 pA / < ± 0.001 nA
Response Time
-
Low range rate: < 12 seconds, High range rate: < 2 seconds
-
Low range charge: <2 seconds, High range charge: <0.3 seconds
Main Software Functions
-
Channel 1 Measurement, Channel 2 Measurement, Channel 1 & 2 Measurement, Scintillator Measurement, Check Source Library, Chamber Library, Scintillator Baseline Library, Preferences
Charge Collection Modes
-
Timed Mode - Selectable timer from 1 to 600 seconds in 1 second increments
-
Continuous Mode (Manual Start/Stop) - Press Start to begin collection and Stop to end
-
Trigger Mode – current exceeding pre-selected trigger values starts and stops collection
Countup/Countdown Resolution
-
1 second (count up for Continuous Mode, count down for Timed Mode)
Trigger Levels for Charge
-
CollectionStart value must be larger than stop value (set via Preferences interface)
-
Start0.2 - 9.9 pA
-
Stop0.1 - 9.8 pA
Operating Parameters
-
Temperature15 to 35 °C
-
Relative Humidity20 to 80% non-condensing
-
Pressure650 to 770 mmHg
Storage Parameters
-
Temperature-15 to 50 °C
-
Relative Humidity10 to 95% non-condensing
-
Pressure600 to 800 mmHg
Display
-
Color LCD TFT touch screen. VGA (640x480) resolution, 6.4” diagonal
Input
-
2 channels, BNC two lug triaxial connector (TNC option is available.)
Bias Voltage
-
± VDC for the following: 0 to 1000 VDC in 1 VDC increments, 350 μA max
Zeroing
-
Zero via button push - Display indicates zeroing in progress
Ports
-
RS232 (Default: Uni-directional 19,200 baud, 8 data bits, no parity, 1 stop bit; Bi-directional operation via user commands, tech note available upon request) (2) USB ports (keyboard, thumb drive, or mouse support) Ethernet (internal use only, not currently supported)
Chamber library calibration coefficients
-
Amp System factorsNsk:μGym²/(h*A)
-
Nsk:Gym²/(h*A)
-
Nx,sk:Rm²/(h*A)
-
NA:GBq/A
-
NA:Ci/A
-
ND,W:mGy/(sec*A)
-
NA:MBq/A
-
NA:mCi/A
-
Coulombs System FactorsNk:Gy/C
-
Nx:R/C
-
NGas:Gy/C
-
ND:Gy/C
-
ND,W:Gy/C
-
Real-Time Clock
-
On board, for data validation and storage
Audio
-
On board speaker for audio feedback on start-up, button presses, operator alerts, etc.
Dimensions
-
Width26.7 cm (10.5 in)
-
Height8.1 cm (3.2 in)
-
Length21.1 cm (8.3 in)
-
Weight2.5 kg (5.45 lbs)
-
Mode of OperationContinuous
Power Requirements
-
100-240 VAC, 0.5 A max, 50/60 Hz input to external power supply, 9 VDC, 2.0 A power supply output to electrometer input, Globtek®, Inc. model GTM 96180-1811-2.0-T.
-
The use of any other power supply and using alternates other than the UL/CSA recognized power cord can degrade minimum safety. The proper replacements from Standard Imaging, Inc. are required for compliance with the requirements of IEC 60601-1.
Product Standards
-
IEC 60601-1:3rd Edition1, IEC 607312
-
1 Externally Certified; 2 Designed to Meet
Shock Classifications
-
Class I - External Power Supply
-
Class II - Electrometer
Publications
Videos