DoseView™ 1D Overview
Easy Operation with Handheld Pendant
- Establish origin position and return to origin with the press of a button.
- Three convenient positioning modes: fast, slow and step (customizable from 0.01 mm to 100 mm).
- One-touch toggle between cm/mm.
Intuitive Data Collection with DV1D Software
- Expedite TG-51 and TRS-398 compliant data collection.
- Remotely adjust settings, conduct measurements and view data in real time.
- Automatic collection of depth curves and single point measurements.
- Automated ratio-based acquisition when using the SuperMAX or two MAX 4000 Electrometers.
Water-Equivalent Bracket Fits Range of Detectors
Adaptable chamber bracket fits most thimble ion chambers and rigid stem parallel plate chambers.
Phantom Available in Two Sizes:
- 34 cm x 30 cm x 36 cm
- 42 cm x 40 cm x 36 cm
1D Scanning Arm Dimensions
-
Height48.26 cm (19.0 in)
-
Width6.77 cm (2.67 in)
-
Length9.44 cm (3.72 in)
Water Tank (REF 91800) Dimensions
-
Height36 cm (14.17 in)
-
Width30 cm (11.81 in)
-
Length34 cm (13.39 in)
Water Tank (REF 91810) Dimensions
-
Height36 cm (14.17 in)
-
Width40 cm (15.75 in)
-
Length42 cm (16.54 in)
Water Tank Material
-
Clear acrylic 0.95 cm (0.375 in)
Weight
-
1D Scanning Arm1.36 kg (3 lbs)
-
Handheld Controller0.23 kg (0.5 lbs)
-
Water Tank (ref 91800, empty)6.35 kg (14 lbs)
-
Water Tank (ref 91810, empty)10.43 kg (23 lbs)
Max Scanning Arm Travel
-
27.5 cm (10.8 in)
Ionization Chamber Holder Diameter Accommodation
-
Max20 mm (0.79 in)
-
Min6 mm (0.24 in)
Backscatter clearance at 25 cm depth
-
˜ 8 cm (3.1) including phantom base
Accuracy of Position
-
± 0.05 mm (0.002 in) over entire 275.00 mm
Repeatability of Position
-
± 0.05 mm (0.002 in) over entire 275.00 mm
Operating Conditions
-
Pressure680 - 800 mm Hg
-
Temperature10 - 40 °C
-
Relative Humidity30 to 75%, non-condensing
Storage Conditions
-
Temperature-40 to 70 °C
-
Relative Humidity0 to 95%, non-condensing
-
Cabling100 ft extension cable provided (standard RS-232 configuration)
Power Requirements
-
AC Output 12 VDC @ 1.25 A
Optional Detector Brackets
-
PTW Markus® (ref 70850)
-
PTW Roos® (ref 70852)
-
Vertical Holder (ref 70851)
Product Standards: CE
Videos
Webinars

More accurate scans in less time with the DoseView 1D--now with new software available! With 0.05 mm accuracy, easily acquire depth-dose measurements in an intuitive, TG-51 and TRS-398 compliant 1D water scanning system.
Oct 26th, 2020 / More Accurate Scans, Less Time: DoseView 1D
More accurate scans in less time with the DoseView 1D. With 0.05 mm accuracy, easily acquire depth-dose measurements Learn about the new software that’s compatible with DoseView 1D and 3D.
Have Questions About the DoseView™ 1D?
Request Pricing for Complete Patient Dosimetry Solution
Patient dosimetry that adapts to daily patient changes. Receive more information about Patient Dosimetry QA Solutions.
Learn more about Standard Imaging Patient Dosimetry